H DAB Color Deconvolution vectors

No Name-4
I have noticed some interesting behavior with the Gabriel Landini's Color
Deconvolution plugin. When I use the 'H DAB' vector set to separate my
images of DAB and hematoxylin stained slides, I get a fair amount of
signal in the 'zero' channel. This is also demonstrated in Landini's
example using the built in 'H DAB' vectors
(http://www.dentistry.bham.ac.uk/landinig/software/cdeconv/cdeconv.html).

I know that I need to measure our actual chromagens separately and then
put these (normalized values) into the user values boxes or add them to
the code. This is also noted by Landini on his page. I am in the process
of doing this now since I know that my chromagens are pretty different
from the default vectors.

The interesting thing is that if I use the default 'H&E DAB' vector set
for Landini's demo image or any of my images, I get what appears to be
much more accurate hematoxylin and DAB channel images. Also, the eosin
channel is very clean compared to the 'zero' channel using the 'H DAB'
vectors. This also holds true when I use these vector sets with my slides
stained only with DAB or only hematoxylin (though the DAB only slide does
show some apparent bleed through into the hematoxylin channel).

The DAB and hematoxylin vector values are the same for both vector sets,
the only difference are the 'zero' vs eosin third vector sets.

I am curious as to why this happens, but more importantly, I am wondering
if I can use the default H&E DAB set as a better rough+ready vector set
(especially for images where I don't have individually stained slides to
determine the ideal vectors.)

I would appreciate any insight into this.

Thanks,
Jim
Reply | Threaded
Open this post in threaded view
|

Re: H DAB Color Deconvolution vectors

Gabriel Landini
On Wednesday 02 November 2005 14:53, No Name wrote:
> I know that I need to measure our actual chromagens separately and then
> put these (normalized values) into the user values boxes or add them to
> the code. This is also noted by Landini on his page. I am in the process
> of doing this now since I know that my chromagens are pretty different
> from the default vectors.

Yes, that is the best: stain the images with only 1 chromogen at a time and
extract the vectors that way.

> The interesting thing is that if I use the default 'H&E DAB' vector set
> for Landini's demo image or any of my images, I get what appears to be
> much more accurate hematoxylin and DAB channel images.

Maybe it looks accurate, but the idea is that one first defines the colours
used to make sure that the unmixing result is what one intends. I have also
noted that DAB does vary across labs but also the haematoxylin is quite
variable (specially since there are several methods).

Make sure that you colour-correct all the images (by dividing the original
image by the brightfield and multiplying the result by 255; you can use the
Calculator_Plus for doing it in one step without having to resort to 32bit
conversion).

> The DAB and hematoxylin vector values are the same for both vector sets,
> the only difference are the 'zero' vs eosin third vector sets.

In the H DAB option the third component is calculated as the orthogonal to the
other 2 available (some shade of green). In the H, E, DAB one the 3rd is the
eosin colour, so the transformation is forced to separate into those 3
instead. (that means that the 2 colour spaces are not the same). I guess that
this difference (by luck?) compensates for any inaccuracies that there may be
in the vectors.

> if I can use the default H&E DAB set as a better rough+ready vector set
> (especially for images where I don't have individually stained slides to
> determine the ideal vectors.)

The problem is, I think, that one is never sure that the other 2 vectors get
separated correctly; they may  have some degree of mixing between them. If
staining is 'exclusive' then you would see the effect, but if you have
co-localisation of the colours it may be difficult to realise the size of the
error, despite that you have an almost empty "3rd image".

If single-stained images are not available, you could try to define the
vectors on areas which are uniquely labelled with each stain, but
unfortunately one cannot make it always sure that this is so.

I hope it helps.

Cheers,

Gabriel
Reply | Threaded
Open this post in threaded view
|

Re: H DAB Color Deconvolution vectors

Smith, Mike
Gabriel,

Your color deconvolution plugin is just what I need, thanks!

Is the Ruifrok and Johnston article available online anywhere, I could only find an abstract.

Mike Smith
Reply | Threaded
Open this post in threaded view
|

Re: H DAB Color Deconvolution vectors

Gabriel Landini
On Thursday 03 November 2005 20:27, Smith, Mike wrote:
> Your color deconvolution plugin is just what I need, thanks!
>
> Is the Ruifrok and Johnston article available online anywhere, I could only
> find an abstract.

Not that I am aware of. Perhaps one can order it through interlibrary loan.

There is a new version coming out today. I will post it to the list when ready
to download.

Cheers,

Gabriel
Reply | Threaded
Open this post in threaded view
|

Re: H DAB Color Deconvolution vectors

No Name-4
In reply to this post by No Name-4
Gabriel,

Thank you very much for your insights into the Color Deconvolution plugin.
I didn't understand how the 3rd vector was created for 2 color stain sets
before. Also, thank you and Dennis for the updated version of Color Decon
- stack capability and the better image names are a big help.

I am still having some trouble getting my DAB and Hem vectors to produce
accurate output. My DAB vector appear to be very close to the built-in DAB
vector but my hematoxylin vector is a bit different (mine:0.778, 0.563,
0.277 vs. Rufroik:0.650, 0.700, 0.290). Using the built in vector set 'H
DAB' I get a result very similar to the example on your web page
(significant signal in the third 'green' vector over both DAB and Hem
stained regions.) Using my vector set for 'H DAB' results in a 'grey'
third channel but now all three resulting images have many pure white
(255) pixels scattered throughout. Does this indicate that the vectors are
too close to each other for accurate separation or just that I still don't
have the correct vectors?

Thanks,
Jim




Gabriel Landini <[hidden email]>
Sent by: ImageJ Interest Group <[hidden email]>
11/02/2005 10:48 AM
Please respond to ImageJ Interest Group

 
        To:     [hidden email]
        cc:     (bcc: James Deeds/PH/Novartis)
        Subject:        Re: H DAB Color Deconvolution vectors


On Wednesday 02 November 2005 14:53, No Name wrote:
> I know that I need to measure our actual chromagens separately and then
> put these (normalized values) into the user values boxes or add them to
> the code. This is also noted by Landini on his page. I am in the process
> of doing this now since I know that my chromagens are pretty different
> from the default vectors.

Yes, that is the best: stain the images with only 1 chromogen at a time
and
extract the vectors that way.

> The interesting thing is that if I use the default 'H&E DAB' vector set
> for Landini's demo image or any of my images, I get what appears to be
> much more accurate hematoxylin and DAB channel images.

Maybe it looks accurate, but the idea is that one first defines the
colours
used to make sure that the unmixing result is what one intends. I have
also
noted that DAB does vary across labs but also the haematoxylin is quite
variable (specially since there are several methods).

Make sure that you colour-correct all the images (by dividing the original

image by the brightfield and multiplying the result by 255; you can use
the
Calculator_Plus for doing it in one step without having to resort to 32bit

conversion).

> The DAB and hematoxylin vector values are the same for both vector sets,
> the only difference are the 'zero' vs eosin third vector sets.

In the H DAB option the third component is calculated as the orthogonal to
the
other 2 available (some shade of green). In the H, E, DAB one the 3rd is
the
eosin colour, so the transformation is forced to separate into those 3
instead. (that means that the 2 colour spaces are not the same). I guess
that
this difference (by luck?) compensates for any inaccuracies that there may
be
in the vectors.

> if I can use the default H&E DAB set as a better rough+ready vector set
> (especially for images where I don't have individually stained slides to
> determine the ideal vectors.)

The problem is, I think, that one is never sure that the other 2 vectors
get
separated correctly; they may  have some degree of mixing between them. If

staining is 'exclusive' then you would see the effect, but if you have
co-localisation of the colours it may be difficult to realise the size of
the
error, despite that you have an almost empty "3rd image".

If single-stained images are not available, you could try to define the
vectors on areas which are uniquely labelled with each stain, but
unfortunately one cannot make it always sure that this is so.

I hope it helps.

Cheers,

Gabriel
Reply | Threaded
Open this post in threaded view
|

Re: H DAB Color Deconvolution vectors

Gabriel Landini
On Friday 04 November 2005 16:23,  [hidden email] wrote:
> I am still having some trouble getting my DAB and Hem vectors to produce
> accurate output. My DAB vector appear to be very close to the built-in DAB
> vector but my hematoxylin vector is a bit different (mine:0.778, 0.563,
> 0.277 vs. Rufroik:0.650, 0.700, 0.290).

You could try to make a image by stitching 2 or 3 images of singly stained
material and note the "ROI vector matrix values" by clicking "show matrices"
for all the stains, repeat this several times and calculate an average value
for each to minimise sampling error). Also try to sample in areas that have
strong stain (but not near black, obviously). Note that the optical density
values from the ROIs are an average, so try to select the ROIs small and as I
suggested above on areas well stained.
Then use "User values" to enter those average values and let the plugin
calculate the translation matrix values (then one should insert those in the
plugin code).

> Using the built in vector set 'H
> DAB' I get a result very similar to the example on your web page
> (significant signal in the third 'green' vector over both DAB and Hem
> stained regions.) Using my vector set for 'H DAB' results in a 'grey'
> third channel but now all three resulting images have many pure white
> (255) pixels scattered throughout.
> Does this indicate that the vectors are
> too close to each other for accurate separation or just that I still don't
> have the correct vectors?

Not sure, it could be a bug.
The 3rd colour, if 0, is defined automatically, so it depends on the other 2.
Please post somewhere an image and a macro with the user values you defined to
see if this is reproducible or if there is a bug in the code.

Please do not email me a large image because until Monday I will be accessing
the net through a slowish modem.

About the white pixels, what kind of image format are you using? Are they
acquired as jpegs? Where one of the stains is not present, the result is
supposed to be 255 or near.

Cheers,

Gabriel
Reply | Threaded
Open this post in threaded view
|

Re: H DAB Color Deconvolution vectors

4N6site.com
In reply to this post by No Name-4
I have programmed a color deconvolution plugin for Photoshop-plugin compatible hosts (Photoshop and other main image editing programs plus a number of free image viewers/editors). I intended it mainly for forensic purposes (color separation and removal) but I think it should be fine for  the purposes discussed here as well (if not, I might add features).
I am giving it away for free via 4N6site.com and you can find examples, instructions etc there as well.

Cheers,
Charles